Product Name:
High Performance Liquid Chromatography Instrument
Ambient Humidity:
20-85% Relative Humidity(non-condensing)
Qualitative Repeatability:
≤0.2%
Quantitative Repeatability:
≤0.4%
Flow Rate Range:
0.0001-10.000mL/min
Max Operating Pressure:
42MPa
Highlight:

Wayeal Liquid Chromatography Instrument

,

220V Liquid Chromatography Instrument

,

Wayeal hplc liquid chromatography

Payment & Shipping Terms

Minimum Order Quantity:
1
Price:
Negotiate
Packaging Details:
Wooden Case
Delivery Time:
30 days
Payment Terms:
L/C, T/T
Supply Ability:
50 Pieces Per Month

Product Description

LC3200 HPLC Liquid chromatography High Performance Liquid Chromatograph

Introduction of High Performance Liquid Chromatography Instrument

High performance liquid chromatograph (HPLC) is an instrument based on the principle of high performance liquid chromatography, which is mainly used to analyze less-volatile and thermal-unstable organic compounds with high boiling point and large molecular weight. It consists of reservoir, pump, sample injector, chromatographic column, detector and recorder.

 

Features of High Performance Liquid Chromatography Instrument

High Pressure Pump

Customized high-precision gradient methods that guarantee extremely high gradient repeatability.
Patented two-stage suspension technology high pressure flow pump, improve the service life of the system
Built-in degasser for both high-pressure pumps and sampler cleaning fluid degassing

Autosampler

Completely self-developed autosampler. The polar coordinate design and various injection methods meet the needs of different customers. Equipped with an auxiliary air pump, there is a cooling mode, and an intelligent lighting mode is optional for easy observation.

Column Oven

Heating, cooling, two-way air bath temperature control, gentle temperature control, uniform heating, high stability.
Provide 2-way 250mm chromatographic column + guard column installation position to meet the needs of most users.

UV Detector

High-precision reflective lens, 1800-line blazed grating, to ensure extremely low stray light and great linear range.

Workstation
With local deployment and network deployment functions, to meet the needs of users of different scales

Technical Data of High Performance Liquid Chromatogaphy Instrument

Manufacturer: Wayeal
Product Name: Liquid Chromatography Instrument
Working Power Supply 220V, 50Hz
Ambient Temperature 10~40°C
Relative Humidity 20~85%
Qualitative Repeatability ≤0.2%
Quantitative Repeatability ≤0.4%
High Pressure Constant Flow Pump  
Flow Rate Range 0.001~10.000mL/min
Setting Step 0.001mL/min
Pressure Range 0~42MPa
Pressure Pulsation 0.1MPa
Flow Rate Set Error ±0.2%
Flow Rate Stability RSD≤0.06%
Gradient Error ±1%
Column Oven  
Temperature Control Range 4~85°C
Temperature Stability ±0.02°C
Temperature Setting Resolution 0.1°C
Autosampler  
Injection Volume Repeatability <0.5%RSD
Linearity >0.9999
Cross Contamination <0.002%
Injection Volume Accuracy Error ±1%
Sample Positions 120
Injection Volume 1~100μL
UV Detector  
Baseline Noise ≤2.0*10-6Au
Baseline Drift ≤1*10-5Au/h
Mini Detection Concentration ≤3.0*10-9g/mL (naphthalene/methanol)
Spectral Range 188~900nm
Wavelength Error ≤±0.1nm
Wavelength Repeatability ≤±0.1nm
Linearity Range ≥103
DA Detector  
Baseline Noise ≤4*10-5Au
Baseline Drift ≤3*10-5Au/h
Mini Detection Concentration ≤2*10-8g/mL
Spectral Range 190~800nm
Wavelength Error ≤±0.1nm
Wavelength Repeatability ≤±0.1nm
Linearity Range ≥ 104
ELS Detector  
Baseline Noise ≤ 0.0037mV
Baseline Drift ≤ 0.0026mV/30min
Evaporation Gas Flow Range (0 ~ 3) SLM
Atomization Temperature Range Room temperature ~ 90℃
Evaporation Temperature Range Room temperature~110℃
Detection Chamber Temperature Range Room temperature~60℃
Input Air Pressure (4 ~ 7) bar
FL Detector  
Baseline Noise ≤ 5×10-6 FU
Baseline Drift ≤ 2 × 10-5 FU/30min
Spectral Range (200 ~ 650) nm
Wavelength Error ± 0.1 nm
Linearity Range ≥ 104
RI Detector  
Detection Range 0.25 ~ 512 μRIU
Linearity Range ≥ 600 μRIU
Response Time 0.1, 0.25, 0.5, 1, 1.5, 2, 3, 6 sec;
Cell Volume 8μL